Cardiorespiratory fitness predicts how long you live as strongly as smoking or diabetes, or more, yet almost nobody has had it measured breath by breath. A wireless mask is trying to change that, with real accuracy and real limits.

Presented By Our Partners

Here is a quiet paradox at the center of longevity science. The study most often quoted to prove that fitness keeps people alive, a 2018 Cleveland Clinic analysis of 122,007 adults published in JAMA Network Open, never measured a single breath of oxygen. It estimated fitness from how long people lasted on a treadmill. Even so, the signal was enormous: people in the lowest fitness group carried roughly five times the risk of death of the elite group over a median of 8.4 years, a gap larger than the one attached to smoking or diabetes in the same patients.

Estimates found that. Measurement is what lets you train it. The VO2 Master Analyzer, made by VO2 Master Health Sensors of Vernon, British Columbia, and tested by researchers under the name VO2 Master Pro, is a portable metabolic analyzer that measures oxygen uptake directly through a face mask and streams it to a phone. As of October 2026 it sells for $6,895 to $7,595 on the company’s store, and it goes wherever an athlete, a soldier, or a patient can walk.

The question is whether a mask with a fuel cell can stand in for a room full of equipment. Four university validation studies now give a surprisingly detailed answer. It depends, more than the marketing suggests, on how you calibrate, how long you test, and what you need the number to do.

The short answer: The VO2 Master Analyzer is a credible, portable way to measure oxygen uptake and ventilation for coaching, field research, and fitness testing. Independent labs have found its oxygen readings anywhere from within 1 percent to about 13 percent of reference values. It measures oxygen only, not carbon dioxide, and it is not a medical device, so it does not replace clinical exercise testing.

  • Best for: coaches, exercise physiologists, gyms, and researchers who need VO2 and ventilation data outside a lab.
  • Not for: clinical cardiopulmonary exercise testing, fuel use (fat versus carbohydrate) analysis, or unsupervised personal use.
  • Price (October 2026): $6,895 to $7,595 per kit on VO2 Master’s store, plus an annual service option and consumables.
  • Regulatory status: not FDA regulated; VO2 Master states it is not a medical device.
  • Evidence: Moderate; four university validation studies with small samples and mixed accuracy results, none of outcomes.

What Is the VO2 Master Analyzer?

The VO2 Master Analyzer is a portable metabolic analyzer made by VO2 Master Health Sensors Inc. that measures oxygen uptake (VO2) and breathing through a mask worn over the nose and mouth. It sends data by Bluetooth to a mobile app in real time. It is a performance and research tool sold to coaches, gyms, and labs, not a consumer wearable or a medical device.

The company was co-founded by Peter O’Brien and Andrew Sellars and has spent roughly a decade on one idea: putting the gas analysis normally done by a cart next to a treadmill into a unit small enough to sit in front of your face. Independent researchers describe the hardware plainly. A 2025 study from the University of Western Ontario lists a differential pressure flow sensor, a galvanic fuel cell oxygen sensor that reads respired oxygen directly at the mouth, and three interchangeable flow meters sized for rest, moderate effort, and hard exercise. Calibration uses room air for the oxygen sensor and a 3 liter syringe for flow.

VO2 Master markets the analyzer for VO2 max testing, resting metabolic rate testing, sports performance, tactical fitness for military and firefighters, and field research. Its own website describes the analyzer as not a medical device. That distinction matters later, when we get to cost, insurance, and who should be reading the results.

How Does the VO2 Master Work?

The VO2 Master measures how much air you breathe and how much oxygen is missing from it on the way out. Multiply ventilation by the drop in oxygen concentration and you get the volume of oxygen your body kept, liter by liter, every breath. Push that number as high as it will go during a hard test and you have VO2 max.

That subtraction at the mouth is the last step in a long relay, and the relay is the reason the number matters. Oxygen makes up about 20.9 percent of the room air you inhale. It crosses the paper-thin walls of the lung’s air sacs into blood, binds to hemoglobin in red cells, rides each contraction of the left ventricle through the arteries, and finally diffuses into working muscle, where mitochondria use it to burn fat and carbohydrate and hand back carbon dioxide. VO2 is the rate at which that entire chain delivers and consumes oxygen. VO2 max is the ceiling: the point where more work no longer buys more oxygen.

Physiologists describe the chain with the Fick principle. Oxygen uptake equals the blood the heart pumps per minute times how much oxygen the tissues strip from each unit of that blood. Which half sets the ceiling? In a classic 2000 review in Medicine and Science in Sports and Exercise, Bassett and Howley laid out three lines of evidence that the limit in healthy exercising people is mostly delivery, not extraction. When oxygen delivery is changed by blood doping, low-oxygen air, or beta-blocker drugs, VO2 max moves with it. When training raises VO2 max, it does so mainly by increasing maximal cardiac output. And a small muscle flooded with blood can consume oxygen at rates far above what the whole body achieves. VO2 max, in other words, is mostly a measurement of the heart as a pump.

That is also why it fades with age in such a revealing way. In the Baltimore Longitudinal Study of Aging, researchers measured peak oxygen uptake repeatedly in 810 healthy adults aged 21 to 87 over a median of 7.9 years. Decline ran 3 to 6 percent per decade in people’s twenties and thirties and accelerated to more than 20 percent per decade in their seventies and beyond (Circulation, 2005). The oxygen delivered per heartbeat fell in step with peak VO2, while maximal heart rate slipped only 4 to 6 percent per decade. The engine does not mainly lose revolutions. It loses the oxygen moved with each beat.

Now picture what the mask has to cope with at the top of that curve. A well-trained athlete near maximal effort can move roughly 250 liters of air a minute, with tidal breaths near 3.8 liters and breathing rates around 65 a minute, figures the 2026 Maastricht validation team drew from the published literature. The VO2 Master handles that range by swapping flow meters: one for rest (5 to 40 liters a minute), one for moderate work (30 to 160), and one for hard exercise (40 to 220). The fuel cell, a small electrochemical cell whose output current tracks oxygen concentration, reads the exhaled gas at the mouth rather than pumping a sample down a tube to a cart. The app turns flow and oxygen into VO2, ventilation, breathing rate, and tidal volume, live.

One thing it does not read is carbon dioxide. In both of the most recent independent studies, the analyzer reported oxygen uptake but not carbon dioxide output, and the 2026 Maastricht team stated it measures only VO2. Without carbon dioxide, there is no respiratory exchange ratio, the ratio of carbon dioxide produced to oxygen consumed that labs use to estimate whether you are burning mostly fat or mostly carbohydrate. For VO2 max and ventilation, that does not matter. For fuel use questions, it is decisive.

How Accurate Is the VO2 Master? What the Research Shows

Independent studies put the VO2 Master’s oxygen readings between about 1 percent and 13 percent from reference values, depending on the study, the software version, and how the device was calibrated and tested. Ventilation readings were strong in the two most recent studies. Oxygen accuracy is good enough for coaching and field research and is less certain for decisions that hinge on a single threshold.

Metabolic analyzers are tested two ways. A metabolic simulator is a machine that “breathes” a precisely known mixture at a known rate, so researchers know the true answer to compare against. Human testing compares the device with an established lab cart on real people, where the truth is less certain but the conditions are real. The VO2 Master has been through both.

The first human comparison. At Alma College in Michigan, Montoye and colleagues tested an early version against a ParvoMedics lab cart (International Journal of Exercise Science, 2020). Eight male cyclists rode an hour-long protocol from 100 to 300 watts twice, and 16 recreationally active men completed three maximal tests each, one with each of two VO2 Master versions and one with the cart. Mean absolute percent differences from the cart were under 9 percent overall for both oxygen uptake and ventilation, and under 12 percent at each stage; the maximal tests came in around 12 percent. Test-retest differences ran 8.9 to 10.9 percent for the VO2 Master against 5.3 to 7.6 percent for the cart. The authors called validity and reliability acceptable for most intensities and the device appealing for field work. The samples were small and all male.

A rough first simulator test. In 2023, a Maastricht University group led by Van Hooren put 15 popular exercise testing systems, including the VO2 Master Pro, on the same metabolic simulator (Scandinavian Journal of Medicine and Science in Sports). By the team’s own later account, the VO2 Master underestimated oxygen uptake by about 11.7 percent, with an absolute error of about 13 percent, a result they described as relatively poor.

A much better result in Ontario. At the University of Western Ontario, Thiessen and colleagues tested two VO2 Master Pro units and two COSMED Quark lab carts on a VacuMed simulator at oxygen uptakes of 1 to 4 liters a minute, with Douglas bag measurements, the old gold standard of collecting exhaled air in a bag, as a further check (Scandinavian Journal of Medicine and Science in Sports, 2025). The VO2 Master’s mean error for oxygen uptake was 0.6 percent, though individual readings ranged from 9.3 percent low to 4.8 percent high. The two units disagreed with each other by as much as 11.0 percent in one direction and 3.6 percent in the other, averaging zero. The authors concluded that, used appropriately in the lab, both systems gave acceptable, reliable data, and they declared no conflicts of interest. They also flagged two practical problems we come back to below.

Featured Partner

Invest in the Infrastructure Behind Modern Medicine

As healthcare expands beyond hospital walls, the buildings and campuses supporting that shift are generating compelling returns for investors who move early. The Healthcare Real Estate Fund offers qualified investors direct access to a curated portfolio of medical office, outpatient, and specialty care facilities.

Learn More →

The Maastricht rematch. In a 2026 follow-up, the same Maastricht team tested five systems, again including the VO2 Master Pro, with no funding and no manufacturer involvement in design or analysis (Scandinavian Journal of Medicine and Science in Sports, 2026). Ventilation was excellent, with an error of 0.61 percent. Oxygen uptake was overestimated by 7.86 percent on the simulator, and the error against the reference in human cycling tests was a closely matching 8.37 percent. Day-to-day technical variation in oxygen readings was about 3 percent. The researchers noted the error had roughly halved since 2023, likely because of software changes including a dedicated simulator mode. They also argued that the Ontario team’s choice to correct for humidity, temperature, and pressure using the simulator’s readings rather than the device’s own would make any analyzer look more accurate than it is in real use. Only four well-trained men completed the human portion.

In patients and in the field. A Duke University team paired the VO2 Master Pro with an impedance heart monitor for a six-minute stepping test in 15 patients with a mean age of 60, comparing it with invasive exercise testing done the same day (JACC: Advances, 2024). Agreement for absolute peak oxygen uptake was good (intraclass correlation 0.77) and weaker for weight-adjusted peak VO2 (0.64). It was a feasibility study, and the authors called for larger validation. At Bond University in Australia, researchers used it to measure the oxygen cost of 5 kilometer marches carrying 30 kilograms in 12 soldiers (Sports, 2025), the kind of outdoor, real-task measurement a cart cannot make.

Read together, the evidence says something more useful than “accurate” or “inaccurate.” Recent ventilation results are very good. Oxygen uptake is good on average and can be very good, but the error has moved with software updates and with the details of calibration and correction. That is the profile of a strong field instrument, not of a reference standard. If you want a framework for weighing claims like these, our explainer on what “clinically validated” actually means is a useful companion.

What the VO2 Master Gets Right

The VO2 Master gets portability right without abandoning measurement. It puts real gas analysis in a mask that works on a road, a trail, or a ruck march, and its ventilation data were excellent in the two most recent independent tests.

It measures instead of guessing. Watches estimate VO2 max from heart rate and pace. A 2022 meta-analysis of 14 validation studies by the INTERLIVE network found that the best of those estimates, ones built on exercise data, had almost no average bias but individual errors spanning roughly 10 mL/kg/min in either direction (Sports Medicine, 2022). That spread can cover the gap between average and excellent fitness. A mask that actually measures oxygen narrows it.

It goes where the work is. The tethered cart forces exercise into the lab. The Bond University load carriage study and the Duke stepping test both used the analyzer in settings a cart handles poorly. For firefighters, soldiers, cyclists on a real climb, and patients doing a bedside test, that freedom is the whole point.

Ventilation is strong. In the 2026 Maastricht data, ventilation error was 0.61 percent, labeled good agreement, and breathing frequency was also within acceptable limits. Ventilation patterns help coaches understand breathing strategy and how hard an athlete is working.

It is improving, and someone is checking. The halving of oxygen error between the 2023 and 2026 Maastricht tests suggests the company iterates. Just as important, independent labs with no funding from the manufacturer keep retesting it. That is the kind of scrutiny few devices at this price get.

It is a realistic price for a testing business. At $6,895 to $7,595 as of October 2026, it is within reach of a coaching business, a university teaching lab, or a physical therapy clinic that wants to add metabolic testing.

Where the VO2 Master Falls Short

The VO2 Master measures oxygen but not carbon dioxide, its oxygen error has ranged widely across studies, it can drift during long steady tests, and it is not a medical device. Those limits matter most for clinical use and fuel use claims.

No carbon dioxide, no fuel story. Because the tested versions report oxygen only, they cannot compute the respiratory exchange ratio, carbohydrate and fat oxidation, or total energy expenditure from gas exchange; the 2026 Maastricht team marked all of those “not available” for this device. If a coach or clinic promises a fat-burning zone from this analyzer, ask how. Earlier versions of this review said the VO2 Master reports carbon dioxide output and respiratory exchange ratio. The published research does not support that for the version tested, and we have removed the claim.

The error is not yet stable. From a 13 percent absolute error in 2023, to a near-zero mean error in Ontario, to an 8 percent overestimate in Maastricht in 2026, the device’s oxygen accuracy has depended on software and method. The Maastricht authors gave a concrete example of why that matters: an error near 8 percent could lead a firefighter to wrongly pass or fail a required VO2 standard, or misclassify a patient’s fitness before major surgery.

It can drift. In a 20-minute steady simulation, both VO2 Master units in the Ontario study showed oxygen readings creeping upward by 110 and 170 mL per minute, apparently from a slow change in the oxygen sensor’s reading; the effect appeared in 5 of 6 repeated trials. For short ramp tests it matters little. For long steady sessions or resting metabolic rate measurements, it is a reason to recheck calibration.

Calibration technique matters. The same team found noisy, nonsensical oxygen values when the device was calibrated by attaching the syringe directly instead of through the recommended clear hose adaptor; the manufacturer suggested the abrupt airflow may “shock” the sensor. That is a usable tip and also a warning: results depend on the operator.

Two units may not agree. Between-device differences in the Ontario study ran as large as 11 percent. A team that tests athletes on two different analyzers should not compare their numbers casually.

Lab-validated, field-marketed. Every validation study we found ran in a lab or clinic, and the Ontario authors warned that wind, temperature, humidity, and altitude could change validity outside the lab. None of the studies tested the app’s training zone or threshold features.

It is not a medical device. VO2 Master says so itself. It is not built or labeled for diagnosing heart or lung disease, and a VO2 max number from it should not substitute for a physician-supervised exercise test when symptoms are involved.

VO2 Master vs KORR CardioCoach vs Cosmed K5 vs a Smartwatch Estimate

The VO2 Master is a true gas analyzer, but in the versions tested it measures oxygen only. KORR’s mixing chamber system and the Cosmed K5 add carbon dioxide, and the K5 is the long-standing research portable. A smartwatch estimates rather than measures, which is fine for trends and poor for individual precision.

FeatureVO2 Master AnalyzerKORR systemCosmed K5Smartwatch VO2 max estimate
What it measuresOxygen and ventilation; no carbon dioxide in versions testedOxygen, carbon dioxide, ventilationOxygen, carbon dioxide, ventilationNothing directly; estimates from heart rate and pace
How it samplesBreath by breath at the mouthMixing chamberBreath by breath or micro mixing chamberAlgorithm
Independent oxygen accuracyAbout 0.6% to 13% mean error across simulator studies4.7% oxygen and 11.9% carbon dioxide error (2026 simulator study)Within about 1% to 4% in mixing chamber mode at high rates (2020 study)Individual errors near 10 mL/kg/min either way (2022 meta-analysis)
Fuel use (fat vs carbohydrate)Not availableReported, but large errors in testingReportedNot available
Best forField VO2 and ventilation testing on a budgetIn-clinic treadmill or bike testingResearch-grade portable testingDaily trends at no extra cost

The KORR figures come from the same 2026 Maastricht study that tested the VO2 Master, where KORR’s mixing chamber system also showed large errors in carbohydrate and fat estimates. Our KORR CardioCoach review covers it in depth. The Cosmed K5 is the step up if you need carbon dioxide and a long publication record; see our Cosmed K5 review. Another portable analyzer included in the 2023 Maastricht comparison is covered in our PNOE metabolic analyzer review. For the wrist estimate, our Garmin Forerunner 965 VO2 max review explains what the watch number can and cannot tell you. To see every device in this category, browse our respiratory and VO2 max device reviews.

How Much Does the VO2 Master Cost, and Is It Covered or HSA/FSA Eligible?

As of October 2026, the VO2 Master Analyzer sells for $6,895 to $7,595 on the company’s online store, depending on the kit. Insurance does not cover it, and because VO2 Master states it is not a medical device, do not count on HSA or FSA reimbursement. Most people will pay per test at a facility that owns one.

Beyond the kit price, budget for upkeep. VO2 Master sells an annual service option on its store, and masks, filters, and a calibration syringe are part of running any analyzer properly; the Ontario study’s calibration used a 3 liter syringe.

If you are an individual, the practical question is what a test costs at a gym, lab, or clinic near you, and whether the operator is trained. Ask which flow meter they use for your test, how and when they calibrated, and whether they will give you the raw VO2 and ventilation data rather than only a fitness score. If your goal is clinical, such as evaluating breathlessness or clearance for surgery, you need a physician-ordered exercise test instead.

Who Should Use the VO2 Master, and Who Should Skip It?

The VO2 Master suits trained operators who need VO2 and ventilation data in the field at a reasonable price. It is the wrong tool for clinical diagnosis, for anyone selling fuel use or fat-burning zone analysis, and for people who want to test themselves at home without training.

Good fits:

  • Endurance and team sport coaches who want measured VO2 max and ventilation on the track, the road, or the field.
  • Tactical and occupational fitness programs testing real tasks such as load carriage.
  • University teaching labs and field researchers who can calibrate carefully and report the device’s known error.
  • Gyms and performance clinics adding metabolic testing, with staff trained in exercise physiology.

Better served by something else:

  • Anyone with chest pain, unexplained breathlessness, or known heart or lung disease: you need a clinician-supervised cardiopulmonary exercise test, not a field mask.
  • Clients who want fat versus carbohydrate fuel analysis: choose a system that measures carbon dioxide.
  • Labs that need a publication-grade reference or high-stakes pass or fail decisions: consider a stationary cart or a K5-class system and validate it yourself.
  • Individuals tracking trends at home: a well-validated watch estimate is free with your device and good enough for direction, if not for precision.

If symptoms are part of the reason you want testing, a cardiologist or pulmonologist can order the right study; the Healthcare Discovery doctor directory lists clinicians by state and specialty.

What This Means for You

VO2 max sits closer to the center of healthy aging than almost any number you can measure. In a 2009 JAMA meta-analysis of 33 studies, each one MET increase in fitness, roughly a 1 kilometer per hour faster running pace, was associated with a 13 percent lower risk of death and a 15 percent lower risk of cardiovascular events. The American Heart Association made the case in 2016 that fitness should be treated as a clinical vital sign, because it reclassifies risk beyond smoking, blood pressure, cholesterol, and diabetes. That is why VO2 max belongs in the larger story we tell in the Four Villains of Health and Longevity: a strong heart and well-supplied muscles push back against the Silent Assassin, cardiovascular disease, and the Energy Arsonist, metabolic dysfunction, and the same AHA statement links low fitness to higher death rates from various cancers, the work of the Rogue Hijacker.

But notice what the evidence does and does not say. The mortality studies mostly estimated fitness from treadmill performance. They show that being fitter is protective, not that measuring it with a mask makes you healthier. The value of a device like the VO2 Master is practical: a real baseline, training intensities set from your own physiology instead of a formula, and a check every few months on whether the work is moving the number.

The work itself lives in the Movement pillar. Because VO2 max is mostly a measure of how much oxygen your heart can deliver, the training that raises it is the training that strengthens the pump: regular, sustained aerobic work plus some genuinely hard intervals. The Baltimore data are the urgency. Decline accelerates with every decade, in active and inactive people alike, so the fitness you build now is the reserve you will draw on at 75. Breath and Sleep support the process through recovery, and Mindset is what gets you back out on the hard days. Measure it if you can. Train it either way.

Frequently Asked Questions

How Much Does a VO2 Master Cost?

As of October 2026, the VO2 Master Analyzer sells for $6,895 to $7,595 on the company’s online store, depending on the kit. Owners should also budget for VO2 Master’s annual service option, replacement masks and filters, and a calibration syringe. Insurance does not cover it, and because the company states it is not a medical device, HSA or FSA reimbursement is unlikely.

How Accurate Is the VO2 Master Compared With a Lab Metabolic Cart?

Independent studies have found its oxygen uptake readings between about 1 percent and 13 percent from reference values. A 2025 Western Ontario study found a mean error of 0.6 percent on a metabolic simulator, while a 2026 Maastricht study found about an 8 percent overestimate. Ventilation readings were accurate in both of those studies. Accuracy depends on software version, calibration technique, and test conditions.

Does the VO2 Master Measure RER or Fat Burning?

Not in the versions tested by independent researchers. Studies published in 2025 and 2026 report that the VO2 Master Pro measures oxygen uptake and ventilation but not carbon dioxide output. Without carbon dioxide, a device cannot calculate the respiratory exchange ratio or estimate how much fat versus carbohydrate you are burning. Choose a system that measures both gases if fuel use matters to you.

Is the VO2 Master FDA Cleared?

No. VO2 Master describes the analyzer as not a medical device, and it is sold for sports performance, tactical fitness, coaching, and research. It is not intended to diagnose heart or lung disease. People with symptoms such as chest pain or unexplained breathlessness should ask a clinician about a supervised cardiopulmonary exercise test instead of a field measurement.

VO2 Master vs Smartwatch VO2 Max: Which Is Better?

For precision, the VO2 Master, because it actually measures oxygen. A 2022 meta-analysis of 14 studies found that even the better smartwatch estimates had individual errors of roughly 10 mL/kg/min in either direction, enough to misplace you by a whole fitness category. Watches are free with the device and useful for tracking trends; a measured test gives you a reliable baseline and training intensities.

Sources

  1. Mandsager K, et al. Association of cardiorespiratory fitness with long-term mortality among adults undergoing exercise treadmill testing. JAMA Network Open, 2018. DOI
  2. Kodama S, et al. Cardiorespiratory fitness as a quantitative predictor of all-cause mortality and cardiovascular events in healthy men and women: a meta-analysis. JAMA, 2009. DOI
  3. Ross R, et al. Importance of assessing cardiorespiratory fitness in clinical practice: a case for fitness as a clinical vital sign. Circulation, 2016. DOI
  4. Bassett DR, Howley ET. Limiting factors for maximum oxygen uptake and determinants of endurance performance. Medicine and Science in Sports and Exercise, 2000. DOI
  5. Fleg JL, et al. Accelerated longitudinal decline of aerobic capacity in healthy older adults. Circulation, 2005. DOI
  6. Montoye AHK, Vondrasek JD, Hancock JB. Validity and reliability of the VO2 Master Pro for oxygen consumption and ventilation assessment. International Journal of Exercise Science, 2020. PubMed
  7. Van Hooren B, Souren T, Bongers BC. Accuracy of respiratory gas variables, substrate, and energy use from 15 CPET systems during simulated and human exercise. Scandinavian Journal of Medicine and Science in Sports, 2023. DOI
  8. Thiessen JS, et al. (Re)assessment of the COSMED Quark CPET and VO2Master Pro systems for measuring pulmonary gas exchange and ventilation. Scandinavian Journal of Medicine and Science in Sports, 2025. DOI
  9. Van Hooren B, et al. Technological and biological reliability, and validity of five different CPET systems during simulated and human exercise. Scandinavian Journal of Medicine and Science in Sports, 2026. DOI
  10. Molinger J, et al. Feasibility of a novel augmented 6-minute incremental step test. JACC: Advances, 2024. DOI
  11. Schram B, et al. Sex differences in the metabolic cost of a military load carriage task: a field based study. Sports, 2025. DOI
  12. Molina-Garcia P, et al. Validity of estimating the maximal oxygen consumption by consumer wearables: a systematic review with meta-analysis and expert statement of the INTERLIVE network. Sports Medicine, 2022. DOI
  13. Winkert K, et al. The COSMED K5 in breath-by-breath and mixing chamber mode at low to high intensities. Medicine and Science in Sports and Exercise, 2020. DOI
  14. VO2 Master Health Sensors. VO2 Master Analyzer store listing (pricing as of October 2026). VO2 Master

VO2 Master Analyzer at a glance

Hardware price
$6,895 to $7,595
Subscription
None required
First year total
$6,895
Form factor
Mask-based portable VO2 system
FDA status
Has FDA clearance
HSA/FSA
Not confirmed

Measures Breathing frequency, RER, VCO2, Ventilatory thresholds, VO2max

Pricing verified October 20, 2026. Maintained in the Healthcare Discovery device database and updated automatically.

Free Daily Briefing

The Latest Longevity Science.
Delivered Every Morning.

Join researchers, physicians, and health professionals getting daily breakthroughs in AI-driven medicine, epigenetics, and longevity research.

Support the research that powers this editorial

No spam. Unsubscribe anytime. We respect your inbox.

Similar Posts