Cosmed K5 Review: Accuracy, Modes, Price, and Who It Suits
A single number drawn from your breath once helped doctors decide who needed a new heart. The Cosmed K5 puts the machine that measures it on your back, and the research shows how close it comes to the lab.
Between October 1986 and December 1989, cardiologists at the Hospital of the University of Pennsylvania did something unusual with every ambulatory patient referred to them for a heart transplant. Before deciding, they had each one exercise to exhaustion while the oxygen in their breath was measured. The result, published in Circulation in 1991, gave cardiologists a threshold they could act on. Of 114 patients, those whose peak oxygen uptake stayed above 14 milliliters per kilogram per minute had one-year survival of 94 percent, as good as survival after transplant. Patients with low values who were turned down for other medical reasons had one-year survival of only 47 percent. Peak oxygen uptake was the best single predictor of who would live.
That measurement is called cardiopulmonary exercise testing, or CPET: exercise while a machine analyzes every breath for both oxygen and carbon dioxide. For decades it meant a cart, a treadmill, and a lab. The Cosmed K5, made by COSMED of Rome, Italy, is the company’s wearable version, a 900 gram unit worn on a harness that does the same gas analysis wherever you can move, and as of October 2026 it is still listed as COSMED’s wearable metabolic system.
We found at least ten validation and reliability studies of it on PubMed, almost all from university labs. The results are reassuring, with one important footnote about which of its two modes to trust when the effort gets hard.
The short answer: The Cosmed K5 is a research and clinical grade wearable metabolic system that measures oxygen, carbon dioxide, and ventilation breath by breath or through a built-in mixing chamber. Independent studies show near lab-level accuracy at low and moderate effort. At high effort, its mixing chamber mode is more accurate than breath by breath. It is institutional equipment sold by quote.
- Best for: university labs, sports science programs, clinical researchers, and hospitals that need full gas exchange data in the lab or the field.
- Not for: individuals, gyms on a budget, or anyone who needs a simple VO2 max number.
- Price (October 2026): COSMED does not publish a price; institutions buy through quotes and distributors.
- Regulatory status: FDA 510(k) cleared as COSMED K5 Wearable Metabolic Technology (K162515).
- Evidence: Strong for measurement accuracy, from many independent validation studies; mode choice matters at high intensity.
What Is the Cosmed K5?
The Cosmed K5 is a wearable metabolic system made by COSMED that measures oxygen uptake, carbon dioxide output, and ventilation during rest or exercise. It is worn on the back with a harness, connected to a face mask, and can store or transmit data. It is used for cardiopulmonary exercise testing, energy expenditure research, and resting metabolic rate measurement.
Independent researchers have described the hardware in detail. A University of Tennessee team reported the unit measures 174 by 111 by 64 millimeters and weighs 900 grams including its battery and oxygen sensor, with a rechargeable lithium ion battery lasting up to four hours, a 3.5 inch display, Bluetooth and USB connections, storage for up to 2,048,000 breaths, and a weather-sealed IP54 case. Oxygen is measured by a galvanic fuel cell, carbon dioxide by a digital infrared analyzer, and airflow by a bidirectional digital turbine with a flow range of 0.08 to 16 liters per second. A University of Rome team added that the K5 includes a 10 Hz GPS receiver, ANT+ wireless sensor support, and an external carbon dioxide scrubber for calibration.
The K5’s defining design choice is what COSMED calls IntelliMET, dual sampling technology. The unit can analyze each breath individually, the approach of its predecessor, the K4b2, or route exhaled gas through a small dynamic mixing chamber driven by a constant flow pump, an approach COSMED borrowed from its Fitmate series. Those two modes behave differently under stress, and that difference turns out to be the most important practical fact about the device.
On regulation, 510(k) database listings of FDA records show the device cleared under the name COSMED K5 Wearable Metabolic Technology, number K162515. That puts it in a different category from the many fitness analyzers sold as non-medical devices.
Can your doctor see this data? Yes, under a clinician's care. Cardiopulmonary exercise testing system run by clinics and labs.
How Does the Cosmed K5 Work?
The K5 measures how much air you move, how much oxygen you remove from it, and how much carbon dioxide you add to it. From those three signals it calculates oxygen uptake, carbon dioxide output, their ratio, and dozens of derived values. The second gas is what makes it a full CPET system rather than an oxygen meter.
Start with the oxygen. Room air is about 20.9 percent oxygen. Each breath carries it into the lungs, across into the blood, and out to muscles on every beat of the heart, where mitochondria burn fuel with it. The K5’s turbine counts the liters flowing through the mask; its fuel cell, a small electrochemical cell whose current rises and falls with oxygen concentration, reads how much oxygen is missing from the exhaled air. Multiply one by the other and you have oxygen uptake. Push it to the limit and you have VO2 max.
Now the carbon dioxide, which tells two more stories. The first is about fuel. Burning fat consumes more oxygen for each molecule of carbon dioxide produced than burning carbohydrate does, so the ratio of carbon dioxide out to oxygen in, the respiratory exchange ratio, drifts upward toward 1 as exercise gets harder and carbohydrate takes over. The second story is about acid. At high intensity, working muscle produces hydrogen ions faster than it can clear them, and the blood buffers them with bicarbonate, a reaction that releases extra carbon dioxide. A 2026 Maastricht validation study built that bicarbonate buffering directly into its simulated breathing patterns, because real lungs show it. The extra carbon dioxide drives breathing harder, and ventilation suddenly climbs faster than oxygen use. That bend in the curve is the ventilatory threshold, and it is visible only when you measure both gases.
Clinicians read more from the same signals. The ratio of ventilation to carbon dioxide output, called the VE/VCO2 slope, shows how efficiently the lungs clear carbon dioxide. In a study of 704 heart failure patients followed for two years, the VE/VCO2 slope was the strongest prognostic marker in every body weight group, and peak VO2 added no further prognostic value once it was known (American Journal of Cardiology, 2007). In another study of 695 stable heart failure patients, a breathing pattern called exercise periodic breathing emerged as the strongest predictor of cardiac death, followed by the VE/VCO2 slope (Journal of Cardiac Failure, 2010). None of that information exists without carbon dioxide analysis.
The two sampling modes handle the gas differently. In breath by breath mode, the K5 samples gas at the mask through a drying line and aligns each gas reading with the flow of that same breath, which captures fast changes such as the first seconds of a sprint. In mixing chamber mode, exhaled gas from several breaths is blended and sampled steadily, which gives more stable concentrations. Think of breath by breath as a high-speed camera and the mixing chamber as a long exposure. One sees motion. The other sees a cleaner picture of a scene that is holding still, or nearly so.
How Accurate Is the Cosmed K5? What the Research Shows
The K5 matches simulators and Douglas bags closely at low to moderate effort, often within about 2 percent for oxygen uptake. Against stationary carts, results vary more, especially at rest and in mixing chamber mode. At high intensity, breath by breath mode underestimates oxygen uptake by several percent, while mixing chamber mode stays closer to reference.
Against a breathing machine. At the University of Rome “Foro Italico,” Guidetti and colleagues ran two K5 units in breath by breath mode against a VacuMed metabolic simulator at 14 metabolic rates from 0.9 to 4 liters of oxygen per minute (PLoS One, 2018). Mean differences were minus 0.04 percent for oxygen uptake, minus 0.50 percent for ventilation, and minus 1.03 percent for carbon dioxide, none statistically significant. The smallest change the system could reliably detect was 3.8 percent for oxygen. The authors stated that COSMED and VacuMed had no role in the study.
Against the Douglas bag. The Douglas bag, which collects exhaled breath in a bag for analysis, is the old criterion method. At the University of Tennessee, Crouter and colleagues compared both K5 modes with Douglas bags in 15 people at rest and while cycling from 50 to 250 watts (PLoS One, 2019). Both modes measured ventilation within 2.1 liters per minute and oxygen uptake within 0.08 liters per minute of the bag. Breath by breath mode underestimated carbon dioxide at 150 watts and above, by 0.12 to 0.31 liters per minute, and both modes gave respiratory exchange ratios 0.03 to 0.08 lower than the bag at 100 watts and above.
Featured Partner
Invest in the Infrastructure Behind Modern Medicine
As healthcare expands beyond hospital walls, the buildings and campuses supporting that shift are generating compelling returns for investors who move early. The Healthcare Real Estate Fund offers qualified investors direct access to a curated portfolio of medical office, outpatient, and specialty care facilities.
Learn More →The high-intensity footnote. At Ulm University in Germany, Winkert and colleagues ran two K5 units simultaneously, one in each mode, against a simulator and Douglas bags in 12 endurance-trained athletes (Medicine and Science in Sports and Exercise, 2020). At low and moderate rates, oxygen uptake was within 0.61 percent to minus 2.05 percent of reference. At high rates, breath by breath mode read 4.63 to 7.27 percent low, while mixing chamber mode read 0.38 to 3.81 percent low. A follow-up in 16 trained men found VO2 max 4.44 percent lower in breath by breath mode than in mixing chamber mode during the same test (International Journal of Sports Physiology and Performance, 2021). The team’s recommendation was direct: test endurance athletes at high workloads in mixing chamber mode.
Against stationary carts. In Gran Canaria, Perez-Suarez and colleagues compared the K5 with a Vyntus CPX lab cart in 16 volunteers (Frontiers in Physiology, 2018). At rest and at 60 watts, breath by breath readings agreed well with the cart. At 130 to 160 watts, breath by breath mode underestimated oxygen uptake by 6.6 percent. Mixing chamber mode overestimated it by 13.4 percent at rest, 5.8 percent at 60 watts, and 4.8 percent at the higher intensity. The authors, who declared no conflicts, still called the K5 almost as accurate as a state-of-the-art stationary cart. Most recently, a Baylor Scott and White team compared it with a ParvoMedics TrueOne 2400 cart in 31 young adults doing Bruce protocol treadmill tests (Proceedings of Baylor University Medical Center, 2026). Test-retest reliability for oxygen and carbon dioxide was excellent, with intraclass correlations of 0.90 to 0.98, though the K5 read slightly high at low intensity and slightly low near peak.
Walking and resting. Slow, steady work is harder for portable systems than it sounds. In 27 adults walking at 1.5 to 4.0 miles per hour, a Syracuse University team found K5 oxygen readings reliable (intraclass correlations of 0.64 to 0.85) and similar to its predecessor, with stronger reliability at faster speeds (European Journal of Applied Physiology, 2020). In 35 walking adults at the University of Granada, the K5 read oxygen uptake, carbon dioxide, and energy expenditure higher than an Omnical cart, with coefficients of variation of 2.6 to 6.3 percent (Medical Engineering and Physics, 2024). For resting metabolic rate in 43 adults aged 18 to 84, a Spanish team found the K5 in mixing chamber mode overestimated most outcomes against an Oxycon Pro and published correction equations (Scandinavian Journal of Medicine and Science in Sports, 2023).
The big comparisons. In the 2023 Maastricht test of 15 exercise testing systems on one simulator, the K5 was among the devices assessed, and the authors concluded that only a few systems showed consistently acceptable error. Because the K5 is so widely trusted, researchers also use it as the reference for other devices: it served as the criterion for testing the PNOE analyzer during walking (2025) and Samsung Galaxy Watch energy estimates in 148 adults (2026).
Taken together, the evidence is among the best for any wearable metabolic system. It also carries one plain instruction: the K5’s mode is a methodological decision, not a default. If you want a framework for weighing validation claims like these, our explainer on what “clinically validated” actually means is a useful companion.
What the Cosmed K5 Gets Right
The K5 gets the full picture right. It measures both gases with lab-level accuracy at most intensities, offers two sampling modes for different questions, and carries a deep, largely independent validation record and an FDA clearance.
Both gases, everywhere. Oxygen alone gives you VO2 max. Oxygen plus carbon dioxide gives you ventilatory thresholds, fuel use, ventilatory efficiency, and the patterns heart failure specialists rely on. The K5 delivers that full panel on a trail, a court, or a hospital corridor.
A mode for each question. Breath by breath captures fast transitions; the mixing chamber steadies hard efforts and long steady states. The Ulm data show why that flexibility matters: in mixing chamber mode, high-intensity oxygen readings were within about 4 percent of reference.
Units agree with themselves. In a separate Ulm study, two identical K5 units fed the same signals showed within-unit technical error below 2 percent on a simulator (PLoS One, 2020). For a lab tracking the same athletes over years, that repeatability is what makes change detectable.
It holds up in the field. In the Gran Canaria study, 14 people completed two 13 kilometer walks at 5 kilometers per hour wearing the K5. Energy expenditure per kilometer repeated with a coefficient of variation of 4.5 percent across more than two hours of continuous work.
Independent scrutiny. Several of the key validation studies state that the manufacturer had no role or that the authors had no conflicts. That is not universal in device research, and it counts.
Where the Cosmed K5 Falls Short
The K5’s breath by breath mode loses accuracy at high intensity, its carbon dioxide and fuel use readings are less precise than its oxygen readings, it reads differently from some carts during walking and rest, and it requires trained operators. Its price is not public.
Breath by breath can underread hard efforts. Underestimates of 4.6 to 7.3 percent at high rates in the Ulm study, and a 4.44 percent lower VO2 max in the follow-up, are large enough to blur an athlete’s real progress over a training block. Labs that test at maximal effort should standardize on the mixing chamber.
Carbon dioxide is the weaker gas. The Tennessee study found carbon dioxide underestimated at higher workloads in breath by breath mode and respiratory exchange ratios lower than the Douglas bag. Because fat and carbohydrate estimates are calculated from that ratio, small gas errors become large fuel errors. Treat a K5 “fat-burning zone” as an estimate.
Unit-to-unit differences in people. The Ulm reliability study found that during real exercise, two identical units in breath by breath mode differed by more than 2 percent for some outcomes. Its authors advised assigning one unit per athlete where possible, and accounting for lower reliability in multicenter studies.
Low intensity and rest are trickier. The Granada and Spanish resting metabolic rate studies both found the K5 reading higher than stationary carts. If your question is resting metabolism, use the published correction equations or a cart built for that measurement.
It is not a self-serve device. Calibration involves room air, reference gas, the flowmeter, the scrubber, and timing delay, done before testing. A K5 in untrained hands is an expensive way to collect bad data.
No public price. COSMED sells the K5 by quote. Earlier versions of this review gave a price range; we could not verify it and have removed it.
Cosmed K5 vs VO2 Master vs KORR vs a Stationary Metabolic Cart
The K5 sits between oxygen-only field devices and stationary lab carts. It measures both gases like a cart, moves like a field device, and has a deeper independent validation record than the other portable systems compared here. A stationary cart remains the reference in the lab, and a field device such as the VO2 Master is lighter on the budget.
| Feature | Cosmed K5 | VO2 Master Analyzer | KORR system | Stationary lab cart (e.g. Vyntus CPX) |
|---|---|---|---|---|
| Gases measured | Oxygen and carbon dioxide | Oxygen only in versions tested | Oxygen and carbon dioxide | Oxygen and carbon dioxide |
| Sampling | Breath by breath or mixing chamber | Breath by breath at the mouth | Mixing chamber | Breath by breath; some carts add mixing chamber |
| Independent oxygen accuracy | Within about 2% in simulator and Douglas bag tests at low to moderate effort; at high effort, mixing chamber within about 4%, breath by breath up to 7% low | About 0.6% to 13% mean error across simulator studies | 4.7% oxygen, 11.9% carbon dioxide error (2026 simulator study) | 3.3% oxygen error for Vyntus CPX (2026 simulator study) |
| Where it works | Lab or field, worn on a harness | Lab or field, worn on the face | Clinic or lab | Lab only |
| Best for | Research and clinical CPET in or out of the lab | Coaching and field VO2 on a budget | In-clinic metabolic testing | Reference measurements and clinical CPET |
The VO2 Master, KORR, and Vyntus figures come from Maastricht’s 2026 simulator study and the other studies in our VO2 Master review. The KORR system is covered in our KORR CardioCoach review, and a Hong Kong study that used the K5 as its reference judged the PNOE comparable during walking; see our PNOE metabolic analyzer review. If your current VO2 max comes from a watch, our Garmin Forerunner 965 VO2 max review explains how that estimate compares with a measured test. For every device in this space, browse our respiratory and VO2 max device reviews.
How Much Does a Cosmed K5 Cost, and Is It Covered or HSA/FSA Eligible?
As of October 2026, COSMED does not publish a price for the K5; universities, hospitals, and teams buy it through quotes from COSMED or its distributors. Individuals do not buy one. They pay for a CPET at a lab or clinic that owns one, and a physician-ordered CPET may be covered by insurance.
Clinical CPET is a billable procedure with its own billing code, CPT 94621, used when a clinician orders the test for a medical reason, such as heart failure evaluation or risk assessment before surgery. Whether your plan covers it, and what you owe, depends on your insurer and your diagnosis, so ask for an estimate before you test. When a clinician orders the test, your out-of-pocket share is generally a qualified medical expense for HSA and FSA purposes; confirm with your plan administrator.
Fitness-only testing at a university or performance lab is usually self-pay. Ask which device and which mode the lab uses for maximal tests, how recently it was calibrated, and whether you will receive your thresholds and the raw data, not just a VO2 max number.
Who Should Use the Cosmed K5, and Who Should Skip It?
The K5 suits institutions that need full, validated gas exchange data in the lab and the field. Individuals should not buy one, but many will benefit from a test on one. Budget-limited gyms and coaches who only need VO2 max are better served by simpler analyzers.
Good fits:
- University exercise physiology and nutrition labs running validation, energy expenditure, or field research.
- Elite sport programs measuring thresholds and fuel use in training environments.
- Hospital and research CPET programs that need a portable, FDA cleared system.
- Clinical trials where peak VO2 or gas exchange is an outcome.
Better served by something else:
- Individuals: book a test instead of buying the device.
- Gyms and coaches who only need VO2 max and ventilation: a field analyzer such as the VO2 Master costs less.
- Resting metabolic rate clinics: a dedicated stationary cart, or published correction equations if you use the K5.
- Anyone with chest pain, fainting, or new breathlessness: you need a physician-supervised clinical CPET, not a performance test.
If you need a cardiologist or pulmonologist to order and interpret clinical CPET, the Healthcare Discovery doctor directory lists clinicians by state and specialty.
What This Means for You
The Cosmed K5 is a reminder of how much medicine can read in a breath. In a 2018 Cleveland Clinic study of 122,007 adults, the least fit had about five times the risk of death of the most fit over a median of 8.4 years (JAMA Network Open). In the Pennsylvania transplant study, peak oxygen uptake sorted heart failure patients into those who could safely wait and those who could not. CPET reads the Silent Assassin, cardiovascular disease, directly through the heart’s ability to deliver oxygen. It also matters for the Rogue Hijacker: in a 2020 meta-analysis of seven studies of esophagectomy, surgery most often done for esophageal cancer, higher preoperative peak VO2 was linked to fewer cardiopulmonary complications and unplanned ICU admissions and to better one-year survival (Annals of Surgical Oncology). Those are the stakes in the Four Villains of Health and Longevity.
The honest limit is that a test does not change the number. Training does. In a 2023 JAMA Network Open meta-analysis of high-intensity interval training before major surgery, eight studies with 627 patients showed peak VO2 rising by about 2.6 mL/kg/min more than with standard care, and eight studies with 770 patients showed fewer postoperative complications (odds ratio 0.44), though the studies varied widely. The Baltimore Longitudinal Study of Aging showed what happens without that work: peak oxygen uptake declines faster with every decade, reaching more than 20 percent per decade after 70 (Circulation, 2005).
So the K5 serves the Movement pillar as a compass. A good test gives you a measured VO2 max, the ventilatory thresholds that set your training intensities, and a baseline to beat. The Breath pillar shows up in the data too: how efficiently you ventilate, and where your breathing breaks from your oxygen use. Measure once, train with purpose for a few months, and measure again. The machine on your back is only the instrument. The change is yours.
Frequently Asked Questions
How Much Does a Cosmed K5 Cost?
As of October 2026, COSMED does not publish a price for the K5. Universities, hospitals, and sports programs buy it through quotes from COSMED or its distributors, and the final price depends on configuration and service. Individuals rarely buy one; instead they pay for a cardiopulmonary exercise test at a lab or clinic that owns the device.
Is the Cosmed K5 Accurate?
Yes, across many independent studies. Against a metabolic simulator, two K5 units measured oxygen uptake within 0.04 percent on average, and against Douglas bags it measured oxygen within 0.08 liters per minute from rest to 250 watts. At high intensity, breath by breath mode read 4.6 to 7.3 percent low, so researchers recommend mixing chamber mode for hard efforts.
Is the Cosmed K5 FDA Cleared?
Yes. FDA 510(k) database listings show the device cleared as COSMED K5 Wearable Metabolic Technology under number K162515. A 510(k) clearance means the FDA found it substantially equivalent to devices already on the market, not that it was tested for every clinical use. It is sold to clinics, hospitals, and research labs rather than consumers.
Breath by Breath or Mixing Chamber: Which K5 Mode Should I Use?
Use breath by breath mode when you need to capture fast changes, such as the first seconds of a sprint or oxygen kinetics. Use mixing chamber mode for maximal tests and steady efforts at high intensity. In a 2020 Ulm University study, mixing chamber mode read oxygen within about 4 percent of reference at high rates, while breath by breath read up to 7.3 percent low.
Cosmed K5 vs VO2 Master: Which Is Better?
The Cosmed K5 is the more complete instrument because it measures carbon dioxide as well as oxygen, which enables ventilatory thresholds, fuel use estimates, and clinical markers such as the VE/VCO2 slope. The VO2 Master lists for $6,895 to $7,595 as of October 2026 and works in the field, but the versions tested measure oxygen only. Choose by the questions you need answered.
Sources
- Mancini DM, et al. Value of peak exercise oxygen consumption for optimal timing of cardiac transplantation in ambulatory patients with heart failure. Circulation, 1991. DOI
- Guidetti L, et al. Validity, reliability and minimum detectable change of COSMED K5 portable gas exchange system in breath-by-breath mode. PLoS One, 2018. DOI
- Perez-Suarez I, et al. Accuracy and precision of the COSMED K5 portable analyser. Frontiers in Physiology, 2018. DOI
- Crouter SE, et al. Accuracy of the Cosmed K5 portable calorimeter. PLoS One, 2019. DOI (doi: 10.1371/journal.pone.0226290)
- Winkert K, et al. The COSMED K5 in breath-by-breath and mixing chamber mode at low to high intensities. Medicine and Science in Sports and Exercise, 2020. DOI (doi: 10.1249/MSS.0000000000002241)
- Winkert K, et al. Inter- and intra-unit reliability of the COSMED K5: implications for multicentric and longitudinal testing. PLoS One, 2020. DOI (doi: 10.1371/journal.pone.0241079)
- Winkert K, et al. Differences in VO2max measurements between breath-by-breath and mixing-chamber mode in the COSMED K5. International Journal of Sports Physiology and Performance, 2021. DOI (doi: 10.1123/ijspp.2020-0634)
- DeBlois JP, White LE, Barreira TV. Reliability and validity of the COSMED K5 portable metabolic system during walking. European Journal of Applied Physiology, 2020. DOI (doi: 10.1007/s00421-020-04514-2)
- Leal-Martin J, et al. Optimization of VO2 and VCO2 outputs for the calculation of resting metabolic rate using a portable indirect calorimeter. Scandinavian Journal of Medicine and Science in Sports, 2023. DOI (doi: 10.1111/sms.14425)
- Alcantara JMA, et al. Validity and between-days reliability of two different metabolic systems for measuring gas exchange during walking. Medical Engineering and Physics, 2024. DOI (doi: 10.1016/j.medengphy.2024.104281)
- Brown KD, et al. Validity and reliability of the K5 wearable metabolic system during high-intensity treadmill testing in healthy, young adults. Proceedings (Baylor University Medical Center), 2026. DOI (doi: 10.1080/08998280.2026.2647611)
- Van Hooren B, Souren T, Bongers BC. Accuracy of respiratory gas variables, substrate, and energy use from 15 CPET systems during simulated and human exercise. Scandinavian Journal of Medicine and Science in Sports, 2023. DOI (doi: 10.1111/sms.14490)
- Van Hooren B, et al. Technological and biological reliability, and validity of five different CPET systems during simulated and human exercise. Scandinavian Journal of Medicine and Science in Sports, 2026. DOI (doi: 10.1111/sms.70184)
- Kwok MMY, et al. Assessing the validity and reliability of the PNOE for measuring cardiometabolic outcomes during walking exercise. Journal of Functional Morphology and Kinesiology, 2025. DOI (doi: 10.3390/jfmk10020159)
- Ferreira ARP, et al. Validity of Galaxy Watch for estimating energy expenditure during intermittent running. JMIR Formative Research, 2026. DOI (doi: 10.2196/83090)
- Chase P, et al. Relation of the prognostic value of ventilatory efficiency to body mass index in patients with heart failure. American Journal of Cardiology, 2007. DOI (doi: 10.1016/j.amjcard.2007.08.042)
- Guazzi M, et al. Development of a cardiopulmonary exercise prognostic score for optimizing risk stratification in heart failure: the PROBE study. Journal of Cardiac Failure, 2010. DOI (doi: 10.1016/j.cardfail.2010.04.014)
- Sivakumar J, et al. The role of cardiopulmonary exercise testing as a risk assessment tool in patients undergoing oesophagectomy: a systematic review and meta-analysis. Annals of Surgical Oncology, 2020. DOI (doi: 10.1245/s10434-020-08638-9)
- Clifford K, et al. Association of preoperative high-intensity interval training with cardiorespiratory fitness and postoperative outcomes among adults undergoing major surgery. JAMA Network Open, 2023. DOI (doi: 10.1001/jamanetworkopen.2023.20527)
- Mandsager K, et al. Association of cardiorespiratory fitness with long-term mortality among adults undergoing exercise treadmill testing. JAMA Network Open, 2018. DOI (doi: 10.1001/jamanetworkopen.2018.3605)
- Fleg JL, et al. Accelerated longitudinal decline of aerobic capacity in healthy older adults. Circulation, 2005. DOI (doi: 10.1161/CIRCULATIONAHA.105.545459)
- U.S. Food and Drug Administration 510(k) database listing K162515, COSMED K5 Wearable Metabolic Technology. 510(k) record
- COSMED. K5 wearable metabolic system product page. COSMED
Cosmed K5 CPET at a glance
- Subscription
- None required
- Form factor
- Portable CPET system (mask-based)
- FDA status
- Has FDA clearance
- HSA/FSA
- Not confirmed
Measures Cardiac output, RER, VCO2, Ventilatory thresholds, VO2max
Pricing verified October 20, 2026. Maintained in the Healthcare Discovery device database and updated automatically.